GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 22
When you perform a calibration, luminance measurement or QA test, PIP MODE will
be turned off automatically. You can test the correct input, using the following:
1. Manually turn PIP MODE OFF on the display via control button.
2. For case (a), (c), (e) and (f), you should only see the image of display area #1,
otherwise changes in connection sequence are required. For case (b) and (d),
change some settings to see the image of display area #2 image as well.
3. Turn ON PIP MODE (PbP) (hereinafter referred to as PbP) and reconfirm the
screen order.
4. Refer to the display's documentation for PIP MODE. The identification of the
display within Windows can be changed by settings of the screen resolution.
Perform procedure 1 to 3 if you changed settings.
Examples of incorrect setup (I.e. changing the settings via the OSD’s SWAP function)
Screen order in Windows setting view and actual display screen are mismatched.
Figure 4: Example of incorrect display setting when using PIP MODE
Screen setting in Windows
(a) Incorrect example 1
Screen setting on display
1
2
Screen setting in Windows
Screen setting on display
2
1
3
1
2
3
2
1
(b) Incorrect example 2
Screen setting in Windows
Screen setting in Windows
Screen setting on display
3
1
4
2
1
2
3
4
Screen setting on display
2
1
3
1
2
3
(d) Incorrect example 3
(f) Incorrect example 4