GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 168
13. MD215MG EDID Serial Number Update Tool
13.1. Overview
The NEC model MD215MG is supported by GammaCompMD QA Client. All communication is
performed via DCC/CI commands over a USB cable. Multiple MD215MG displays which are
connected to the same system must be calibrated sequentially. The serial number of each
unit is required to identify a MD215MG display. Normally it is stored in the EDID data of a
display and retrieved and stored in the Windows registry during a system start.
In case of the model MD215MG, the serial number is not stored within in the EDID, but in a
different storage location inside the display. This chapter instructs on how to read the serial
number from the MD215MG and to store it in the GammaCompMD QA Client database,
using the MD215MG EDID Serial Number Update Tool.
Operating System Environment
Windows XP 32/64, Windows 7 32/64bit, Windows 8 / 8.1 32/64bit
Software Environment
GammaCompMD QA Client Version 4.0.40 or later
Hardware Environment
One or more MD215MG with connected USB cable(s) directly connected to the system.
NOTE:
- A daisy chain setup of the USB cables from display to display is not supported.
- USB 3.0 connections are not supported. Please use a USB 2.0 connector.
- Using this tool is a one-time action, as long as the display / display controller configuration
is not changed.