NEC GammaCompMD QA Uživatelský manuál Strana 119

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 182
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 118
GammaCompMD QA Client User Manual
Copyright © NEC Display Solutions Ltd., 2006 -2015 119
FIT Test Error Define the maximum permissible range for the slope of the
straight line that was calculated when executing a linear regression
analysis for the JND-Index interval data, which is back calculated
from the luminance characteristics after grayscale calibration. This
can be set in a range from 0.0001 to 0.9900.
Send to client workstation
To display a dialog box on the selected display when the preset
limit is exceeded during measurement, check this checkbox.
Send to Server Check this box to send an alert to the GammaCompMD QA Server
when the preset limit is exceeded during measurement.
Select All/Deselect All Select or deselect all tick boxes for both Send to client
workstation and Send to Server.
Conformance (2) tab
Figure 102: Alert Setup Conformance (2) dialog box
GSDF Error (If the GSDF value exceeds the preset limit, activate the alert)
Define the reset limit when the conformance test is performed. If
User defined value is selected, the preset limit can be selected or
entered.
Zobrazit stránku 118
1 2 ... 114 115 116 117 118 119 120 121 122 123 124 ... 181 182

Komentáře k této Příručce

Žádné komentáře