Figure 118: White Luminance Measurement Start screen
Figure 119: White Luminance Measurement Report dialog box
7.7.3.2. Black Luminance Measurement
Click the Black Luminance Measurement in Administrator to display the Black Luminance
Measurement dialog box, as shown in Figure 120.
If both display sensors and external sensors have been set for various displays, measurement
starts with the displays with display sensors, then continues with displays for which external
sensors are set. Although measurement is executed simultaneously for multiple display
sensors (except MD212G3/MD215MG/MD211G5 and MD302C6), it is executed in ascending
order of display numbers for external sensors.